dc.description.abstractÇalışmamızda, radyoterapide doz dağılımını modifiye etmek için kullanılan fiziksel ve sanal kama filtrelerin dozimetrik özelliklerini inceleyerek, uygulamadaki avantaj ve dezavantajlarının saptanması amaçlandı. Fiziksel ve sanal kama filtrelerin kama faktörleri (WF), yüzde derin doz (%DD) eğrileri, yüzey ve çevre dozların verilerini karşılaştırmak amacıyla elde edildi. Ölçümler Siemens Artiste cihazında 6 MV ve 15 MV X-Işın enerjileri kullanılarak 5x5, 10x10, 15x15 ve 20x20 cm2 alanlarda, SSD: 100 cm’de su ve katı su fantomunda farklı iyon odaları ile yapıldı. Kama faktör ölçümleri 5 cm derinlikte Farmer iyon odası ile yapılırken, %DD eğrileri Semifleks iyon odası ve yüzey dozu ölçümlerinde ise Markus paralel plak iyon odası kullanıldı. Çevre dozu ölçümleri için 2-D Array kullanılarak dmax ve 5 cm derinliklerde alındı. Aynı derinlikte, küçük alanlardan büyüğe doğru gidildikçe WF’leri fiziksel ve sanal kama için artığı görüldü. Sanal kama filtrelerin çevre dozu fiziksel kama filtreye göre yüksek bulundu. Sanal kama için küçük alanlarda kama açısının çevre dozlarına etkisinin olmadığı ancak büyük alanlarda kama açısıyla birlikte etkinin arttığı görüldü. Fiziksel kama için derinlik ve alan boyutu arttıkça çevre dozunun arttığını buna karşın kama açısı arttıkça çevre dozunun azaldığını saptandı. Sonuç olarak sanal kama filtreler set-up kolaylığı ve tedavi süresinin kısalması sebebiyle tercih edilebilir. Ayrıca daha fazla açı alternatifi sunması planlamada avantaj sağlar.tr_TR
dc.description.abstractThe aim of this study is to determine the advantages and disadvantages of physical and virtual wedge filters used in radiothreapy for modifing dose distribution by investigating the dosimetric features of them. In order to comparing the wedge factors (WF), percent depth dose (%DD) curves, surface and the environment dose values were obtained. Tha measuements were performed in Siemens Artiste lineer accelerator at the energies of 6 MV and 15 MV with the fields of 5x5, 10x10, 15x15 ve 20x20 cm2 at SSD 100 cm by using different ion chambers in water and solid water fantoms. The WF measurements were done at 5 cm depth by Farmer ion chamber, while the percent DD curves and surface dose values were measured by Semifleks ion chamber and Marcus paralel plate ion chamber, respectively. The environment dose measurements using the 2-D Array at dmax and 5 cm depth were performed. It was found that the physical and the virtual WFs were increased from small to large fields in the same depth. The environment dose values of the virtual wedge was higher than the physical wedge. For the virtual wedge, the effect of wedge angle on environment dose was negligable unlike large fields. For the physical wedge, the environment dose was deacreased by increasing wedge angle and increased by increasing depth and field size. As a result, virtual wedge filters would be preferred due to the ease of setup and shortening of treatment time. Additionally virtual wedge filters were provided more angle alternatives compared to physical wedge filters.en_US
dc.identifier.citationŞahin, A. vd. (2016). “Yüksek enerjili fotonlarda fiziksel ve sanal kama filtrelerin dozimetrik özelliklerinin karşılaştırılması”. Uludağ Üniversitesi Tıp Fakültesi Dergisi, 42(1), 1-7.tr_TR
dc.identifier.endpage7tr_TR
dc.identifier.issn1300-414X
dc.identifier.issue1tr_TR
dc.identifier.startpage1tr_TR
dc.identifier.urihttps://dergipark.org.tr/tr/download/article-file/421396
dc.identifier.urihttp://hdl.handle.net/11452/18962
dc.identifier.volume42tr_TR
dc.language.isotrtr_TR
dc.publisherUludağ Üniversitesitr_TR
dc.relation.journalUludağ Üniversitesi Tıp Fakültesi Dergisi / Journal of Uludag University Medical Facultytr_TR
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergitr_TR
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectBilgisayarlı tedavi planlamatr_TR
dc.subjectYüksek enerjili X-Işınıtr_TR
dc.subjectKonformal radyoterapitr_TR
dc.subjectSanal kama filtretr_TR
dc.subjectFiziksel kama filtretr_TR
dc.subjectDozimetritr_TR
dc.subjectRadiotherapy treatment planningen_US
dc.subjectHigh energy X-Rayen_US
dc.subjectConformal radiotherapyen_US
dc.subjectVirtual wedge filteren_US
dc.subjectPhysical wedge filteren_US
dc.subjectDosimetryen_US
dc.titleYüksek enerjili fotonlarda fiziksel ve sanal kama filtrelerin dozimetrik özelliklerinin karşılaştırılmasıtr_TR
dc.title.alternativeComparing dosimetric properties of physical and virtual wedge filters at high energy photonsen_US
dc.typeArticleen_US
dc.typeAraştırma makalesitr_TR

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
42_1_1.pdf
Size:
805.9 KB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: