Publication:
Dosimetric control of dose distribution calculated in computerized treatment planning system

No Thumbnail Available

Date

2011-01-01

Authors

GÜRLER, ORHAN
Şahin, Sevim
Gürler, Orhan
Gözcü, Sema
Kurt, Meral
Şengül, Kansu
Altay, Ali
Kahraman Çetintaş, Sibel
Özkan, Lütfi

Authors

Journal Title

Journal ISSN

Volume Title

Publisher

Kare Publ

Research Projects

Organizational Units

Journal Issue

Abstract

OBJECTIVESThe purpose of this study is providing dosimetric control by comparing dose values calculated in computerized treatment planning system (CTPS) and measured by Linear 2D-Array and ion chambers of various volumes.METHODSFields measured by Farmer, Semiflex and PinPoint ion chambers were created in CTPS. Using Linear 2D-Array, measurements were done in open, isocentric, oblique and wedge filtered fields which created in CTPS. Dose values were compared.RESULTSPinPoint ion chamber measured top dose values whereas Farmer ion chamber measured low values. The most compatible results with CTPS were obtained by Semiflex ion chamber. However, in 2D-Array measurements, concordance was observed with CTPS in central axis, difference reached 20 % in build-up and penumbra regions.CONCLUSIONThe most compatible results with CTPS were obtained by Semiflex ion chamber. Reasons of discordance in 2D-Array measurements were found as; varible dose around build-up region, and not to obtain the exact superpose of profiles' penumbra region.

Description

Keywords

Computerized treatment planning system, Linear 2d-array, Quality assurence in radiotherapy, Science & technology, Life sciences & biomedicine, Oncology

Citation

0

Views

0

Downloads

Search on Google Scholar