Simultaneous determination of metronidazole and miconazole in pharmaceutical dosage forms by RP-HPLC

No Thumbnail Available

Date

2002-11

Journal Title

Journal ISSN

Volume Title

Publisher

Elsevier Science

Abstract

A reversed-phase high performance liquid chromatography (RP-HPLC) method with UV detection is described for the simultaneous determination of metronidazole and miconazole in pharmaceutical dosage forms. Chromatography was carried out on a Cl 8 reversed-phase column, using a mixture of methanol-water (40 + 60, v/v) as a mobile phase, at a flow rate of 1.0 ml min(-1). Sulfamethoxazole was used as an internal standard and detection was performed using a diode array detector at 254 nm. The method produced linear responses in the concentration ranges 10-70 and 1-20 mug ml(-1) with detection limits 0.33 and 0.27 mug ml(-1) for metronidazole and micanozole, respectively. This procedure was found to be convenient and reproducible for analysis of these drugs in ovule dosage forms. (C) 2002 Editions scientifiques et medicales Elsevier SAS. All rights reserved.

Description

Keywords

Pharmaceutical dosage forms, Pharmacology & pharmacy, Reversed-phase high performance liquid chromatography, Amperometric detection, Metronidazole and miconazole, Performance liquid-chromatography, Derivatıve spectrophotometry, Creams, Plasma, Metabolites, Extraction, Nitrate, Drugs

Citation

Akay, C. vd. (2002). "Simultaneous determination of metronidazole and miconazole in pharmaceutical dosage forms by RP-HPLC". Farmaco, 57(11), 953-957.

Collections